Medical Air Systems Inc., can supply complete medical gas equipment
Medical Air Systems Inc., can supply complete medical gas equipment

General Design
Specifications

 

The Medical Compressed Air system is designed for superior performance and quality. The air systems are designed to meet or exceed NFPA 99 recommended guidelines for medical air systems. All systems are piped and wired for single point connection. The standard system includes a control panel complete with full-size NEMA rated starters and alternating controls. Systems are available in simplex and all multiplex arrangements. The air systems are available in the following standard configurations


Tank Mounted on a horizontal receiver.

Base Mounted with vertical receiver.

Stack Mounted with vertical receiver.

Individual compressor modules with vertical receiver.

Medical Air Systems Inc., supply air compressors of the highest quality

 

The air compressors are an air-cooled, oil-less design for supply of clean, high quality air to the medical/surgical air system. Each compressor utilizes grease packed, sealed-for-life bearings and PTFE piston rings to provide reliable operation without conventional lubrication of crankcase and cylinder walls. Low lift, stainless steel valves and anodized valve plates with aerodynamic valve ports ensure durability and long life. All models operate at 72-78dba. An OSHA-compliant belt guard encloses the V-belt drive. Each compressor discharge is equipped with a shutoff valve and a High Air Temperature Shutdown switch wired to the system control panel. Each system control panel includes factory mounted hour meters to monitor factory recommended service intervals as well as running lights to indicate compressor in operation.

Each compressor is driven by a ball bearing, 40° C rise, standard design induction motor with 1.15 service factor.

The system electrical controls are mounted in a NEMA 12 control cabinet, UL Listed as a system. The standard controls include the following features:

Elements

Combination type panel including circuit breaker disconnects, magnetic across the line motor starter switch with three phase overload protection and under voltage release.

An automatic alternator to cycle lead compressor with each use.

Hand-Off-Automatic selector switches.

Lag Compressor In-Use indicator light with horn and connection for remote annunciation.

High-Air-Temperature shutdown indicator lights with horn and connection
for remote annunciation.

Hour meters to monitor factory recommended service intervals.


The system includes properly sized inlet and discharge flex connectors, and vibration isolation pads for field installation. As an option, the system can be supplied with properly sized intake filters with minimum 10 micron filtration for either rooftop or inline mounting.

The system includes one ASME rated air receiver with 4-1/2" pressure gauge, safety relief valve, automatic and manual tank drain, sight gauge glass and 3-valve by-pass.